|
Results and
Discussion
Spectrometer calibration:
The unloaded cavity resonance
was found to be at a frequency of 9.35 GHz. For reasons yet
to be investigated, the cavity resonance frequency did not agree
with calculations for the simple cylindrical cavity. The resonance
frequency found was nearest to the TE112 mode line. The cavity
was not an ideal cavity, so it was expected that the resonance frequency
would be lower, but not by the amount measured. It is interesting
to note that when the cavity was modeled using concentric cylinders,
figure 7, and by using the mode chart in figure 8 to determine the
value of the Bessel function, the calculated resonance frequency
was 9.36 GHz, which strongly agrees with the measured value. This
indicates that the cavity mimics attributes similar to a coaxial
cavity.

Figure 7: Geometric Parameter for a concentric cylindrical
cavity.

Figure 8: TE modes in concentric cylindrical cavities

Figure 9: Loaded Cavity Resonance
The spectrometer was tested
and calibrated using a sample of 2,2-diphenyl 1-picrylhydrazyl
(DPPH). DPPH was used because of its narrow, but intense resonance
line resulting from the free electron associated with one of the
nitrogen atoms. The DPPH was placed in the cavity and tuned
into resonance. The resonance dip is shown in figure 9, the
quality factor (Q) was calculated to be 1870. The absorption resonance
due to the unpaired electrons in the DPPH is shown in figure 10. The
spectrum was observed at room temperature, 298 K, using X-Band microwaves
with a frequency of 9.325 GHz. The magnetic field was swept
at a rate of 100 G per minute through a range of 1000 G. The
magnetic field was modulated at 20 KHz and the signal was filtered
through a phase sensitive lock-in amplifier. The resulting
signals observed were the absorption resonance obtained directly
through the detector, figure 10(a), and the derivative of the absorption
resonance obtained through the lock-in amplifier, figure 10(b). The
absorption resonance was observed at a magnetic field strength of
3325 G. Using the fundamental equation of EPR presented in
the theoretical section (eq-7), a g-value of 2.0035 is obtained. This
agrees with the g-value for DPPH (g = 2.0036 ± 0.0003) to
within the experimental error of the accepted value. The g-value
measured for DPPH slightly deviates from the g-value of the free
electron (g = 2.0023), this indicates that the orbital motion of
the nitrogen bound electron is nearly canceled out by the molecular
structure of the DPPH. Although the g-value peaks at a field
strength of 3325 G, the integrated spectrum reveals that the electrons
resonate over a range of magnetic field strengths. This build
up and decay of the resonance is due to the magnetic interactions
between the electrons and their environment. The number of
DPPH molecules in the sample does not allow all of the electrons
to transition at one specific energy. Further measurement of
the integrated intensity, the area under the signal from the detector,
figure 10(a), allows for the calculation of the concentration of
EPR active species within the sample.

Figure 10: Signal observed
for DPPH through the detector (bottom) and lock-in amplifier (top)
Synthetic Ruby Experiment:
Our sample was cylindrical
in shape with height of 1.3335 cm and a diameter of 0.3193 cm. The
color is deep red which suggests impurities of iron as well as chromium. The
sample is suspended from a small cut of salt crystal and is adhered
with thick grease. It lies inside a glass chamber that protrudes
into the cavity. The synthetic ruby sample and the salt crystal
are connected to a brass cylinder which is in turn connected to
the dewer. The dewer rests on a steel counter and is free to
rotate 360 degrees.
We begin testing the ruby by
making sure our cavity is resonating as well as possible. This
is done by adjusting the waveguide screws and/or the reflector intensity
until the absorption dip in out peak displayed on our oscilloscope
is as close to the center of the peak, and as close to zero as possible.
The magnet is then activated
and swept for short increments of about 1-2.5 min. We then
adjust the sensitivity of the lock-in amplifier until the signal
is sufficient. If the ruby derivative peaks are at an appreciable
height, approximately 1mV, and the derivative signal looks close
to the theoretical derivative signal, then we increase the sweep
time to ten minutes with a range on 10000 G. Since we do not
know where to define zero, we set the dewer to an arbitrary zero
and take data for ten minute sweeps at ten degree increments.
A Hall probe is attached to
the electromagnet and reads off an induced voltage reading in millivolts. A
reading of the voltage value from the probe is recorded at each
peak. It is crucial to record the position of at least one
peak and recommended to record at least two. These will be
used later to correlate the peak positions when the time steps are
converted to the field strength in Gauss. This recording of
the peaks is necessary because the sweep time is never exactly ten
minutes.

Figure 11: Lattice Orientation
Our ruby is axially mirrored
across a 90 degree range and repetitive across a 180 degree range. This
is demonstrated in the isofrequency plot created from laboratory
data. Our absorption peaks are believed to be caused by electron
spin flips within the Cr+3 because our experiments sensitivity
level is too low to resolve anything else. We believe that
the c-axis of our crystal is likely tilted with respect to the z
axis of the dewer that it is hanging from and being rotated with. We
conjecture that the only way for the lattice reaction to change,
if the lattice is repetitive, is for it to be tilted at different
angles to the B field as it is rotated.

Figure 12: detail of Lattice.
As the crystal rotates, different
surfaces are visible to the magnetic field. In
other words, the magnetic flux through the crystalline lattice will
be more or less at a glancing angle, depending on the angle of rotation. This
causes the peaks to appear at different Gauss readings for different
angles within a 90 degree rotation

Figure 13: Isofrequency plot for various orientations of synthetic
ruby
Error Analysis:
There were a significant amount
of errors involved throughout the duration of this experiment. Beginning
with the electromagnet, there were difficulties from the start. When
calibrating the magnet, we used a Hall probe was itself calibrated
with an uncertain value. The value of the calibration magnet
was determined using a current deflection test, which had human
error involved. The magnet was then used to calibrate the Hall
probe, which was then used to read the strength of the electromagnet
to determine the error in the value displayed by the electromagnet.
Error also occurred in the
measurement of the klystron frequency because the frequency meter
used was switched out midway through the experiment and afterwards
the peak value did not match up to the previous one. Currently,
we have no explanation for this phenomenon, but obviously, it had
an effect on the accuracy of our results.
While recording the data for
the peaks in the ruby sample, we took down the voltage values by
hand, which created a great source of human error. Some of
the peak values were not recorded directly in the middle of the
peak and frequently happened slightly after they occurred. This
significantly reduced the accuracy of each peak value and since
there was not sufficient time to take several sweeps of the same
value, no average value could be obtained.
|